| Model Number |
60007 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
09/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.Manufacturer reference #: (b)(4).
|
| |
|
Event or Problem Description
|
|
A site reported to rxsight that a light adjustable lens+ (lal+, sn (b)(6), +20.5d) was explanted from a patient's right eye on (b)(6) 2024, approximately 6 months after completion of light treatments, due to patient dissatisfaction with distance vision.Another lal+ (sn (b)(6), +21.0d) was implanted as a replacement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is submitted to provide the results of the investigation.Updates to sections g3, g6, h2 and h6.The explanted lal (sn (b)(6)) was cut into small fragments during explantation and returned to rxsight.Due to the condition of the lens, an evaluation could not be performed.
|
| |
|