• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS NEEDLE SFTYGLD 23X1 RB; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS NEEDLE SFTYGLD 23X1 RB; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Catalog Number 305902
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 09/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B.3.The date received by manufacturer has been used for this field.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
Material #: 305902.Batch#: 4087139.It was reported that the needle sftygld 23x1 rb safety mechanism broke.The following information was provided by the initial reporter: verbatim: complaint received via email(s) attached.Rcc received a complaint via phone.(b)(6).Ref: 305902.Lot: 4087139.Reporter said multiple clinics have reported issues issue: 2 needle sticks occurred with 2 different mas when medical asst.Were administering vaccines to pediatric pts.Product has a safety device that you activate with your finger and it malfunctioned.In addition customer is reporting that the needle is thinner and weak clinicians are not confident in the retraction function.Pt harm: yes diagnostic testing did occur as a result of the needle sticks.Representative samples of the reported lot are available for evaluation.Sample return kit requested.
 
Additional Manufacturer Narrative
(b)(4) - supplemental mdr - safety mechanism broke (safetyglide) since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.We would be very interested in examining product that does not meet your expectations and our quality standards.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.Examination of the product involved may provide clarification as to the cause for the reported failure.We appreciate you taking the time to bring this observation to our attention.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEEDLE SFTYGLD 23X1 RB
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key20435240
Report Number1213809-2024-00703
Device Sequence Number7627060
Product Code FMI
UDI-Device Identifier30382903059028
UDI-Public(01)30382903059028(11)240327(17)290228(10)4087139
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K951254
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number305902
Device Lot Number4087139
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2024
Supplement Date Received by Manufacturer12/04/2024
Initial Report FDA Received Date10/12/2024
Supplement Report FDA Received Date12/09/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-