| Catalog Number |
205552000 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Illegible Information (4050)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the head trials do not fit onto trial stick.The sizes are not visible on the head trials.These were found after the case so there were no adverse consequences and the patient was not affected.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.Added: g1.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # :(b)(4).Investigation summary = > it was reported that the head trials do not fit onto trial stick.The sizes are not visible on the head trials.These were found after the case so there were no adverse consequences and the patient was not affected.¿ the product was returned to j&j medtech orthopaedics for evaluation.¿ visual inspection of the returned device found it was heavily scratched.The allegation of illegible etch was not confirmed as the investigation did not find sufficient evidence to confirm the reported event.¿ the observed condition of the device is suspected to be attributed to an excessive forceful insertion/extraction technique during trialing or using alternative methods for extraction (such as prying or gripping the trial with pliers or clamps to extract) which are not recommended forms of extraction in the surgical technique.No material or manufacturing defects were observed that would have contributed to the material becoming damaged more easily.¿ a dimensional inspection was not performed since it was not applicable to the complaint condition.¿ a functional evaluation was performed using a sample s/c trial handle angled and it revealed the s/c & mod cath trl 52/28 was able to assemble to the handle successfully.The complaint condition was not replicated.¿ the overall complaint was not confirmed as the observed condition of the s/c & mod cath trl 52/28 would have not contributed to the complained device issue.¿ ¿ based on the investigation findings, the potential cause is traced to unintended use, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot = > a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (ae0403) provided is not a valid finished goods lot# h11 additional narrative: corrected: h3.
|
| |
|
Search Alerts/Recalls
|