MEDOS INTERNATIONAL SARL 4.75 HEALIX ADVANCE KNTLS BR; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 222581
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 09/13/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: a1: patient identifier: (b)(6).E1: the initial reporter's name, facility name and address were not provided.H4: the device manufacture date is currently unavailable as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
This is report 1 of 3 for (b)(4).It was reported through a post market study that a patient had undergone a rotator cuff repair procedure with biceps tenodesis and supra + subscapularis repair on (b)(6) 2024, using two 4.75 healix advance kntls br devices and one super qa+ o/c ds cp-2 device.According to the report, a post operative fall occurred resulting in anatomic neck fracture of the proximal humerus and a retear of the rotator cuff.It was reported that this had ultimately appeared to have healed but a reverse shoulder arthroplasty procedure was needed and performed on (b)(6) 2024.There was no delay in the procedure reported.The status of the patient was unknown.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: h4: the device manufacture date has been updated accordingly.Investigation summary the complaint device was not returned to j&j medtech orthopaedics, therefore unavailable for a physical evaluation.Investigation summary according to the information received, it was reported that postop fall-anatomic neck fracture of the proximal humerus and retore her rotator cuff repair.Ultimately this appears to have healed but a reverse shoulder arthroplasty is needed.The complaint device was not returned to j&j medtech orthopaedics, therefore unavailable for a physical evaluation.This complaint file was opened to document complaints derived through clinical research data.The clinical research data indicated a j&j medtech orthopaedics product failure(s).Since there was no contact information, no follow-up attempts could be performed.It is unknown if complaints derived from this clinical research data were previously reported and documented in the j&j medtech orthopaedics complaint system at the time of occurrence as no product code/lot number information was provided to perform the search.A manufacturing record evaluation was performed for the finished device lot number (203l160), and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.75 HEALIX ADVANCE KNTLS BR
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key20440574
Report Number1221934-2024-03656
Device Sequence Number18209488
Product Code MAI
UDI-Device Identifier10886705030309
UDI-Public(01)10886705030309(10)203L160(17)260630
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K171114
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Catalogue Number222581
Device Lot Number203L160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2024
Supplement Date Received by Manufacturer10/16/2024
Initial Report FDA Received Date10/14/2024
Supplement Report FDA Received Date10/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR