• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC SMART PORT CT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC SMART PORT CT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number CT66LTPD
Medical Device Problem Codes Fracture (1260); Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
Date of Event 09/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2013, plaintiff underwent placement of an implantable vascular access device at (b)(6), where she was implanted with an angiodynamics smartport port, ref # ct66ltpd, lot # 587154.The device malfunctioned, and on (b)(6) 2022, plaintiff underwent surgery to remove the smartport.However, after multiple attempts, the port could not be removed.On (b)(6) 2023, plaintiff underwent another procedure for removal of the device.The device had fractured and the fragmented remains could not be fully removed.As a result of having the smartport implanted, plaintiff has, allegedly, experienced significant mental and physical pain and suffering, has sustained permanent injury, permanent and substantial physical deformity, has undergone and will undergo corrective surgery or surgeries, has suffered financial or economic loss, including, but not limited to, obligations for medical services and expenses, and present and future lost wages.
 
Additional Manufacturer Narrative
The reported device is not available to be returned to the manufacturer for evaluation.An investigation into the root cause for the event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
The customer's reported complaint description of catheter tubing became fractured cannot be confirmed.No port/catheter tubing device was returned for evaluation.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.Device history record review of the packaging/assembly/component lots revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Labeling review: the instructions for use, (16608224-01) which is supplied to the user with this item number, contains the following statements: - absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.- if the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Contraindications · catheter insertion in the subclavian vein medial to the border of the first rib, an area which is associated with higher rates for pinch-off.Potential complications: catheter fragmentation catheter pinch-off drug extravasation (leakage).Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.The following statements reference making the catheter connection to the port housing: - insert catheter into sheath.Position the distal end of the catheter at the desired location.Peel away sheath while withdrawing it from vessel.Care should be taken not to withdraw catheter as sheath is removed.Catheter position should be confirmed radiographically.Secure catheter in place.- trim proximal end of catheter and advance through subcutaneous tunnel to the port pocket.Attach catheter to the port body.- slide the blue strain relief mechanism over the end of the catheter.The tapered end of the blue strain relief mechanism should point away from the proximal end of the catheter.For optimal results, the proximal end of the catheter should be dry.Slide the trimmed end of the catheter tip onto the stem until the catheter is flush with the stem flanges.Slide the strain relief mechanism over the catheter and onto the stem until it contacts the port body.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART PORT CT
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key20442027
Report Number1317056-2024-00177
Device Sequence Number18184035
Product Code LJT
UDI-Device IdentifierH787CT66LTPD0
UDI-PublicH787CT66LTPD0
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K101017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2015
Device Model NumberCT66LTPD
Device Lot Number587154
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2024
Supplement Date Received by Manufacturer11/14/2024
Initial Report FDA Received Date10/14/2024
Supplement Report FDA Received Date11/20/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient SexFemale
-
-