| Catalog Number |
J683G |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The device upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.Additional information was requested, and the following was obtained via: 1.Were there any patient consequences? had to use a 16g needle (extra poke) 2.How long was the delay? minutes but irritating 3.Were there any unexpected outcomes or complications as a result of the prolonged surgery time? no 4.Was there any change in the patient¿s post-operative care due to the prolonged procedure? no 5.Status of product return: picked up from grh by rep d4: udi: the manufacture and expiration dates are currently not available.Therefore, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure in (b)(6) 2024 and suture was used.During the procedure, had 9 boxes total of item j683g from the same lot# ubmpde where all of the sutures have been breaking off of the needle once the needle was pulled through patients skin.The impact has been procedure delay and having to use an additional 16g needle to feed the suture in order to complete the suture line.There were no adverse consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigation summary: the product was returned to ethicon for evaluation.Eight boxes with twelve unopened samples each and one open box with five unopened samples were received for analysis.Product code j683g.As per the sampling plan, a visual inspection was performed on thirty-two samples, the packets were opened.The swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.The functional test was performed using instron equipment and t he pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|