• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71938-01
Medical Device Problem Codes Failure to Power Up (1476); Unable to Obtain Readings (1516)
Health Effect - Clinical Codes Hyperglycemia (1905); Loss of consciousness (2418)
Date of Event 09/20/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product has been requested back for investigation and the customer agreed to return the device; however, no product has been received at this time despite follow up attempts by adc.An investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the freestyle libre reader were reviewed and the dhrs showed the freestyle libre reader passed all tests prior to release.In addition, the review determined that the correct cable and adapter were part of the reader kit pack and there was no indication that the product did not meet specifications.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.If product is returned, a physical investigation will be performed per adc's established processes and procedures and a report will be submitted upon completion of investigation.The date of event is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A battery/no power issue was reported with the abbott diabetes care (adc) device.The customer was unable to test due to the device not powering on with button press or test strip insertion.The customer experienced loss of consciousness and was unable to self-treat.The customer had contact with a healthcare professional where an unspecified capillary blood glucose result was obtained.The customer was administered insulin injection (dose/type unspecified) by the healthcare professional.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14 DAY
Common Device Name
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key20445376
Report Number2954323-2024-40511
Device Sequence Number7995408
Product Code PZE
UDI-Device Identifier00357599002008
UDI-Public(01)00357599002008(10)01A008D
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P160030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number71938-01
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2024
Initial Report FDA Received Date10/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/25/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexMale
-
-