• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KLS MARTIN SE & CO. KG LEVEL ONE CMF; PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KLS MARTIN SE & CO. KG LEVEL ONE CMF; PLATE Back to Search Results
Model Number 50-735-24-09
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 10/02/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation was successfully completed.The results of the investigation have identified no manufacturing issues, no design issues and/or corrective actions necessary at this time.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Event or Problem Description
It was reported cmf plates were identified on imaging as bent due to unknown reasons.They were removed and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
PLATE
Manufacturer (Section D)
KLS MARTIN SE & CO. KG
kolbinger strasse 10
muehlheim/donau, deu 78570
GM  78570
Manufacturer (Section G)
KLS MARTIN SE & CO. KG
kolbinger strasse 10
muehlheim/donau, deu 78570
GM   78570
Manufacturer Contact
melissa bachorski
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key20450353
Report Number9610905-2024-00060
Device Sequence Number17903576
Product Code MQN
UDI-Device Identifier00888118077707
UDI-Public(01)00888118077707
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K032442
Exemption Number2017029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number50-735-24-09
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2024
Initial Report FDA Received Date10/15/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age38 YR
Patient SexMale
-
-