This report is being supplemented to provide additional information based on the device evaluation and legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the customer's complaint was confirmed.Based on the investigation results, it is presumed that the tissue pad was severely worn out by ultrasonic output with the gripping part closed (including after the tissue was cut) without grasping the tissue.As a result of checking the instruction manual, the following description related to this event was found.This event is warned by the instruction manual.Content of the instruction manual was confirmed.The instruction manual contains the following descriptions, and it warns against this event.Instruction manual figure number/revision number: rc2058 10.Drawing number and revision number of ifu: rc2058 10.·do not activate output while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.·do not close the gripping part and output when nothing is grasped between the grasping part and the probe tip, or when it is not possible to confirm whether the living tissue or blood vessel being grasped has been incised.Due to abnormal heat generation caused by friction between the gripping part and the probe tip, the gripping part and the probe tip may be damaged, deformed, or falling off, and part of the tissue pad may peel off, leading to severe wear.When cutting or vessel sealing is performed, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.When treating living tissues or blood vessels, make an incision with light tension so that the incision can be seen.Also, when it is cut off, stop the output immediately.Otherwise, abnormal heat generation due to friction between the tissue pad and the probe tip may lead to damage, deformation, or disconnection of the gripping part (both the stainless-steel part and the fluoropolymer part) and the probe tip, or a part of the tissue pad may peel off.Olympus will continue to monitor field performance for this device.
|