• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TUSKER MEDICAL INC. TULA KIT - BILATERAL; TYMPANOSTOMY TUBE DELIVERY PRODUCT WITH DRUG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TUSKER MEDICAL INC. TULA KIT - BILATERAL; TYMPANOSTOMY TUBE DELIVERY PRODUCT WITH DRUG Back to Search Results
Catalog Number TULAKIT2
Medical Device Problem Code Therapeutic or Diagnostic Output Failure (3023)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: internal complaint reference number: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication reported that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that during a tympanostomy, the patient's ear was thoroughly cleaned prior the iontophoresis with the tula kit.The iontophoresis was setup and completed without any complications.As the doctor made contact with the tympanic membrane the patient screamed in pain and would not let the doctor have a second try, therefore, the parents opted to finish the procedure in the or.The parent made a comment afterward asking if it was lidocaine that was used to numb the area because the patient was mostly non-responsive to lidocaine and recently had a procedure with lidocaine injections and the lidocaine injections were largely ineffective.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TULA KIT - BILATERAL
Common Device Name
TYMPANOSTOMY TUBE DELIVERY PRODUCT WITH DRUG
Manufacturer (Section D)
TUSKER MEDICAL INC.
155 jefferson drive
menlo park CA 94025
Manufacturer (Section G)
TUSKER MEDICAL INC.
155 jefferson drive
menlo park CA 94025
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key20450911
Report Number3012130335-2024-00010
Device Sequence Number7110262
Product Code QJA
UDI-Device Identifier00840128600023
UDI-Public840128600023
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P190016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/27/2024
Device Catalogue NumberTULAKIT2
Device Lot Number23022301
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2024
Initial Report FDA Received Date10/15/2024
Date Device Manufactured02/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient SexUnknown
-
-