• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US MODULAR CATHCART BALL 54MM OD; HEMI HIP IMPLANT : HIP ACETABULAR METAL HEMI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US MODULAR CATHCART BALL 54MM OD; HEMI HIP IMPLANT : HIP ACETABULAR METAL HEMI Back to Search Results
Catalog Number 136354000
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 09/07/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had a right hip dislocation.Revised to total hip on (b)(6) 2024.There was loosening of the femoral stem at the non-cemented interface.No issues during surgery.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: b5.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information was received and states that: was there any surgical delay related to the event? no.Was there any allegation against the augment? no allegations.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR CATHCART BALL 54MM OD
Common Device Name
HEMI HIP IMPLANT : HIP ACETABULAR METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20452750
Report Number1818910-2024-21683
Device Sequence Number12533191
Product Code LZY
UDI-Device Identifier10603295032861
UDI-Public(01)10603295032861(17)331130(10)M51H02
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K903084
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number136354000
Device Lot NumberM51H02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/03/2024
Supplement Date Received by Manufacturer10/15/2024
10/25/2024
Initial Report FDA Received Date10/15/2024
Supplement Report FDA Received Date10/22/2024
10/29/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TAPERED SPACER ARTICUL/EZE +0.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexMale
-
-