| Model Number |
8300AB21 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Insufficient Device Problem Information (3190); Device Stenosis (4066)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/25/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 21mm 8300ab intuity valve was placed under evaluation for valve-in-valve intervention after an implant duration of six (6) years, two (2) months due to unknown reasons.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b1, b2, b4, b5, g3, g6, h2, h6.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 21mm 8300ab intuity valve underwent valve-in-valve intervention after an implant duration of six (6) years, three (3) months due to aortic stenosis and regurgitation.The tavr procedure was performed with a 23mm non-edwards transcatheter valve.The patient left the procedure in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.A definitive root cause is unable to be determined, however, patient and/or procedural factors likely caused or contributed.
|
| |
|