| Model Number |
KNOTLESS T-ROPE SYN-DESMOSIS REPR TI |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an ankle joint surgery the tightrope tore.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.Visual inspection identified that the tightrope of the knotless tightrope® syndesmosis repair implant, titanium was cut.The round button was not returned.Functional testing was not performed due to the damage to the device.The most likely cause of the reported failure can be attributed to user error of the device due to either excessive force used during suture passing or the device coming in contact with another device during use.
|
| |
|
Search Alerts/Recalls
|