• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROLINE SURGICAL INC RENEW MICROSCISSOR TIP, DISPOSABLE; MANUAL DETACHABLE SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROLINE SURGICAL INC RENEW MICROSCISSOR TIP, DISPOSABLE; MANUAL DETACHABLE SURGICAL INSTRUMENT Back to Search Results
Model Number 3122
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
At this time, microline surgical, inc is awaiting a disposable tip of the same lot back from the importer/distributor as the tip involved in this incident was discarded.An investigation will be conducted with the other tip, and a final adverse event report will be submitted with the results.
 
Event or Problem Description
The heatshrink came off during the surgery.The heat shrink was removed from the abdominal cavity, so there was not health problem.The tip was discarded since it was used for the infectious disease patients.
 
Additional Manufacturer Narrative
In response to the recent inquiries from amco incorporated, pertaining to heat shrink falling off of microline surgical inc.'s renew disposable tips (cat#: 3112), we have received the complaints.We apologize for any inconvenience that this issue may have caused.Microline surgical inc.Has initiated a voluntary recall on all parts of affected lot: 00172110 (cat#: 3122) and issued a corrective action (car-0200) to address this problem.Other immediate containment carried out by microline surgical inc.Included inspecting all grasper tips in process and in our stockroom for this defect.Additionally, our quality assurance team performed an independent verification of heat shrink prior to release for every lot of tips being manufactured on the affected assembly line until the car investigation was complete.The root cause identified through the investigation of car-0200 was that the tips were not processed in accordance with the work instruction.The parts were not subjected to elevated temperature in an oven as the assembly process calls for.Therefore, the heat shrink did not firmly attach to the tip assemblies.When investigating further the following factors were identified as contributing to the process failure: the state of work of material was not properly identified during break times and the end of the day.The work area had no visible staging area to identify status of parts (awaiting oven versus already processed through oven).Insufficient inspection and training of operators.The oven process log does not include the number of tips processed.There is no method to reconcile if all parts were placed in the oven.For the implementation stage of car-0200 the assembly line work area was rearranged to provide a separate holding area for tips processed through the heat shrink oven.Product is held in this area until it is 100% inspected and released for further processing.Our assembly work instruction was updated to include more detail on how to perform a tactile pull/push test which will ensure the heat shrink is secure.Our inspection procedure was revised to include a step on visually examining the shrunk tubing.A step was also added to our packaging work instruction as another level of verifying that the heat shrink is firmly attached.Lastly, the oven process log used in tip assembly was revised to record the quantity of tips processing through the oven.All operators have been trained to the updated procedures.Car-0200 is currently in the verification stage where we will ensure the improvements implemented are still being practiced and we will monitor customer complaints for this defect.
 
Event or Problem Description
The heatshrink came off during the surgery.The heat shrink was removed from the abdominal cavity, so there was not health problem.The tip was discarded since it was used for the infectious disease patients.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEW MICROSCISSOR TIP, DISPOSABLE
Common Device Name
MANUAL DETACHABLE SURGICAL INSTRUMENT
Manufacturer (Section D)
MICROLINE SURGICAL INC
50 dunham rd suite 1500
beverly MA 01915
Manufacturer (Section G)
MICROLINE SURGICAL INC
50 dunham rd suite 1500
beverly MA 01915
Manufacturer Contact
scott marchand davis
50 dunham rd suite 1500
beverly, MA 01915
9789229810
MDR Report Key20455207
Report Number1223422-2024-00036
Device Sequence Number7110382
Product Code GEI
UDI-Device Identifier00811099010036
UDI-Public00811099010036
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K962119
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3122
Device Catalogue Number3122
Device Lot Number00172110
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2024
Supplement Date Received by Manufacturer10/15/2024
Initial Report FDA Received Date10/15/2024
Supplement Report FDA Received Date05/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-