• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION MOSAIC; REPLACEMENT HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION MOSAIC; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 305C25
Medical Device Problem Codes Calcified (1077); Intravalvular regurgitation (4068)
Health Effect - Clinical Code Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 09/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information that 16 years and 6 months post implant of this aortic bioprosthetic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement.The reason for evaluation was not reported.It was stated that tortuosity was seen in iliac arteries.No intervention or adverse patient effects were reported.Subsequently, 6 days later, a transcatheter valve was implanted valve-in-valve.The reason for intervention was reported as severe aortic regurgitation.It was also reported that degenerative calcification was present which restricted the movement of 2 of the leaflets.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIC
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key20457199
Report Number2025587-2024-05744
Device Sequence Number7718855
Product Code DYE
UDI-Device Identifier00721902694752
UDI-Public00721902694752
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P990064
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/12/2012
Device Model Number305C25
Device Catalogue Number305C25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2024
Initial Report FDA Received Date10/15/2024
Date Device Manufactured01/16/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient Age73 YR
Patient SexMale
-
-