| Model Number |
P3201D |
| Medical Device Problem Code |
Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No further information is available on the repair of the bed at this time.The investigation is ongoing, however if any additional relevant information is identified following completion of the investigation, the additional relevant information will be submitted in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that sparks are coming from the near side of the head caster.The bed was located at the account.There was no patient/user injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The baxter technician found the power cord needed to be replaced.Per the baxter service manual, perform annual preventive maintenance procedures to make sure all versacare bed components are functioning as originally designed.Examine the plug for damage.Make sure the plug is a one-piece molded plug assembly.If it is not, replace the plug cord assembly.Replace any plug cord assembly that shows any of these: discoloration of the plug molding, around the plug blades.Any signs of cracking., loose fit of the plug blade (the plug blade moves in the molding), verify that the strain relief p-clip is present.Replace the power cord, if damaged.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the power cord to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Event or Problem Description
|
|
It was reported that sparks are coming from the near side of the head caster.The bed was located at the account.There was no patient/user injury reported.
|
| |
|