| Catalog Number |
0279351103 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the black insulation around the shaft in front of the ablation area has partially detached.Replacement was available.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be: 1) probe used as a tool for mechanical displacement of tissue or bone, or to pry or scrub, 2) excessive force used when inserting of removing probe, probe inserted into obstructed passageway, or any forceful impact to the tip of the probe with rigid objects.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|