MAUDE Adverse Event Report: OSTEOPORE INTERNATIONAL PTE LTD OSTEOMESH; BONE FILLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSTEOPORE INTERNATIONAL PTE LTD OSTEOMESH; BONE FILLER Back to Search Results
Model Number PC12(39,10,1)
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Swelling/ Edema (4577)
Date of Event 10/16/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Device had been implanted for a period of approximately 9 months before it was reported on (b)(6) 2023 that patient experienced swelling, soreness, pain and fluid accumulation.On (b)(6) 2024, revision surgery was performed to remove the device.During removal of device, the surgeon reported that device had showed signs of integration with surrounding soft tissue.Additionally, with use of endoscope, a small bump could be observed in the right nostril, none observed in the left nostril.Colourless fluid observed after cutting into bump.No clear signs indicating presence of infection reported by the surgeon.Currently, the patient is recovering well with no more complaints of pain.This complaint is associated with the second of 2 devices implanted in this patient.Refer to medwatch 3007303685-2024-00001 for the first device implanted in this patient.
 
Additional Manufacturer Narrative
This complaint was reported outside of the us market.This device model number is marketed in the us.The device was explanted on (b)(6) 2023 and returned to osteopore international for pathology evaluation.The internal investigation into device production lot included a review of the reported information, review of the device history and complaint history records, and analysis of the returned biopsy specimens.Review of device history record(s) showed no issues during the manufacture of this device.No non-conformity was identified in the review of the following records: production record, sterilization record, certification of irradiation.Histological investigation did not reveal any features of implant rejection and, although no bacteria were observed, an infective aetiology may have been possible.In our investigations, there had been no other complaints reported against the same production lot.Additionally, review of complaint history records from year 2020 revealed no other complaints reported against similar device model number.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSTEOMESH
Common Device Name
BONE FILLER
Manufacturer (Section D)
OSTEOPORE INTERNATIONAL PTE LTD
2 tukang innovation grove
#09-06 jtc medtech hub
singapore, singapore 61830 5
SN  618305
Manufacturer (Section G)
OSTEOPORE INTERNATIONAL PTE LTD
2 tukang innovation grove
#09-06 jtc medtech hub
singapore, singapore 61830 5
SN   618305
Manufacturer Contact
ammar hassanbhai
2 tukang innovation grove
#09-06 jtc medtech hub
singapore, singapore 61830-5
SN   618305
MDR Report Key20459677
Report Number3007303685-2024-00002
Device Sequence Number7718467
Product Code GXP
UDI-Device Identifier08885013200577
UDI-Public(01)08885013200577
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K051093
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model NumberPC12(39,10,1)
Device Lot Number22-0185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/16/2023
Initial Report FDA Received Date10/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age35 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-