MAUDE Adverse Event Report: KEYSTONE DENTAL, INC. IHMAX; I-HEXMRT IMPLANT 7.0 X 11MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KEYSTONE DENTAL, INC. IHMAX; I-HEXMRT IMPLANT 7.0 X 11MM Back to Search Results
Model Number 15866K
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); Osteopenia/ Osteoporosis (2651)
Date of Event 08/29/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Loss of osseointegration.
 
Additional Manufacturer Narrative
Keystone dental inc.Provides labeling with all implants.Failure to osseointegrate and loss of osseointegration (implant mobility) is identified as an adverse reaction and identified in the list of warnings in all endosseous dental implant labeling.The specific cause for a particular complaint is often not readily identified as there are various factors that contribute to the risk of implant failure.These include patient factors such as prior oral infection, poor bone quality or quantity, systemic conditions such as diabetes and uncontrolled hypertension.Technique factors such as overheating of the implant site may cause necrosis of the bone, preventing growth (pelayo et al., 2008; turkyilmaz & al., 2008).Patient habits such as tobacco use (heitz-mayfield & huynh-ba, 2009), alcohol or drug abuse, poor oral hygiene, and bruxism (salvi & braegger, 2009) may also lead to implant failure.In addition, improper surgical technique can lead to implant failure and/or loss of supporting bone (salvi & braegger, 2009).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IHMAX
Common Device Name
I-HEXMRT IMPLANT 7.0 X 11MM
Manufacturer (Section D)
KEYSTONE DENTAL, INC.
154 middlesex turnpike
burlington MA 01803
Manufacturer (Section G)
KEYSTONE DENTAL, INC.
154 middlesex turnpike
burlington MA 01803
Manufacturer Contact
erik roman
154 middlesex turnpike
burlington, MA 01803
7813283490
MDR Report Key20460351
Report Number3005990499-2024-71939
Device Sequence Number12711296
Product Code DZE
UDI-Device IdentifierD76815866K0
UDI-Public+D76815866K0/$$3261028WO0189955
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 10/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number15866K
Device Catalogue Number15866K
Device Lot NumberWO-018995
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2024
Initial Report FDA Received Date10/16/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexMale
-
-