Model Number CT80STPD-VI
Medical Device Problem Codes Obstruction of Flow (2423); Detachment of Device or Device Component (2907); Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device is not available to be returned to the manufacturer for evaluation.An investigation into the root cause for the event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
On or about (b)(6) 2017, plaintiff underwent placement of an angiodynamics smartport product, reference number ct80stpd-vi, lot number 5197392.The device was implanted by dr.(b)(6) , m.D., at (b)(6) hospital in louisville, kentucky, for the purpose of ongoing vein access.On or about (b)(6) 2018, plaintiff presented to (b)(6) hospital in louisville, kentucky, to undergo fragment retrieval and catheter replacement.Plaintiff's catheter had fractured, migrated to the right ventricle, and was also occluded.The fragment was successfully removed, and plaintiff's catheter was replaced.
 
Additional Manufacturer Narrative
The customer's reported complaint description of catheter tubing became fractured and detached cannot be confirmed.No port/catheter tubing device was returned for evaluation.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.Device history record review of the packaging/assembly/component lots revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Labeling review: the instructions for use, (16608102-01) which is supplied to the user with this item number, contains the following statements: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.If the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Contraindications: catheter insertion in the subclavian vein medial to the border of the first rib, an area which is associated with higher rates for pinch-off.Potential complications: catheter fragmentation; catheter pinch-off; drug extravasation (leakage).Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.The following statements reference making the catheter connection to the port housing: insert catheter into sheath.Position the distal end of the catheter at the desired location.Peel away sheath while withdrawing it from vessel.Care should be taken not to withdraw catheter as sheath is removed.Catheter position should be confirmed radiographically.Secure catheter in place.Trim proximal end of catheter and advance through subcutaneous tunnel to the port pocket.Attach catheter to the port body.Slide the blue strain relief mechanism over the end of the catheter.The tapered end of the blue strain relief mechanism should point away from the proximal end of the catheter.For optimal results, the proximal end of the catheter should be dry.Slide the trimmed end of the catheter tip onto the stem until the catheter is flush with the stem flanges.Slide the strain relief mechanism over the catheter and onto the stem until it contacts the port body.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART PORT CT
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key20461396
Report Number1317056-2024-00179
Device Sequence Number7718509
Product Code LJT
UDI-Device IdentifierH787CT80STPDVI1
UDI-PublicH787CT80STPDVI1
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K062414
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2020
Device Model NumberCT80STPD-VI
Device Lot Number5197392
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2024
Supplement Date Received by Manufacturer11/14/2024
Initial Report FDA Received Date10/16/2024
Supplement Report FDA Received Date11/20/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient SexFemale
-
-