Catalog Number MWM012
Medical Device Problem Code Residue After Decontamination (2325)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that upon kit inspection, four bloody/rusted instruments were found.
 
Additional Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Additional Manufacturer Narrative
The reported event was confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the devices were returned on 05th nov 2024 for evaluation.In some of the received disk sizer corrosion marks/residue was observed.After cleaning as per stryker guidelines most of the surfaces especially the metal component were found to be without any residue.However, a few of sizers had black residue stuck between meeting surface of metal and plastic component.These sizers also showed normal signs of wear and abrasion.The source of the residue could not be confirmed.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the available images and above investigation, the issue is most likely user related.It is also not conveyed whether any pre-cleaning was done which is required as per ifu.A real source of contamination could not be confirmed due to lack of further information, but a probable source could be an uncleaned cleaning & sterilization machine at the facility leading to cross contamination.If more information is provided, the case will be reassessed.
 
Event or Problem Description
It was reported that upon kit inspection, four bloody/rusted instruments were found.
 
Additional Manufacturer Narrative
Correction: this event was initially reported in error.
 
Event or Problem Description
It was reported that upon kit inspection, four bloody/rusted instruments were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORM SIZER DISK SIZE 2
Common Device Name
TEMPLATE
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key20461533
Report Number0001649390-2024-00601
Device Sequence Number15667427
Product Code HWT
UDI-Device Identifier00846832082711
UDI-Public00846832082711
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberMWM012
Device Lot NumberCV0322025
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/18/2024
Supplement Date Received by Manufacturer12/11/2024
01/30/2025
Initial Report FDA Received Date10/16/2024
Supplement Report FDA Received Date01/03/2025
01/31/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-