| Model Number |
PM20212 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported they opened a set of rts defib electrodes and the gel was peeling off.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated the product was released accomplishing all quality standards.A photo was provided and visually inspected; the reported issue was observed.No abnormal conditions were found in our process that could trigger this issue.All information received will be used for tracking and trending purposes and we will continue to monitor.
|
| |
|
Additional Manufacturer Narrative
|
|
Sections h3 and h6 evaluation codes were updated.Updated investigation summary: the device history record was reviewed and indicated the product was released accomplishing all quality standards.A photo was provided and a retained device was tested; the reported issue was observed.No abnormal conditions were found in our process that could trigger this issue.All information received will be used for tracking and trending purposes and we will continue to monitor.Sections h3 and h6 evaluation codes were updated.
|
| |
|
Search Alerts/Recalls
|