| Model Number |
39467-125 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Intermittent Loss of Power (4016)
|
| Health Effect - Clinical Codes |
Vascular Dissection (3160); Pericardial Effusion (3271)
|
| Date of Event |
08/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
E1 initial reporter email: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the burr became stuck in lesion, guidewire detachment occurred, requiring additional intervention.The target lesion was located in the severely calcified right coronary artery.A 1.25mm rotapro and rotawire were selected for use.During the procedure, after reaching the lesion, rotablation was performed but was ineffective as burr became stuck in lesion.The burr was removed as a system, mother-in-child system technique, but the rotawire breaks downstream of the burr.At the end of the procedure, coronary dissection at the level of the occlusion with a 2 cm pericardial effusion was noted and pericardiocentesis was performed.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 describe event or problem: updated.H6 device code: updated.E1 initial reporter email: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the burr became stuck in lesion, guidewire detachment occurred, requiring additional intervention.The target lesion was located in the severely calcified right coronary artery.A 1.25mm rotapro and rotawire were selected for use.During the procedure, after reaching the lesion, rotablation was performed but was ineffective as burr became stuck in lesion.The burr was removed as a system, mother-in-child system technique, but the rotawire breaks downstream of the burr.At the end of the procedure, coronary dissection at the level of the occlusion with a 2 cm pericardial effusion was noted and pericardiocentesis was performed.It was further reported that burr stall after a few seconds from the start of rotation.
|
| |
|
Search Alerts/Recalls
|