• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL WAVEONE GOLD SMALL 3-FILE STER 31MM; FILE, PULP CANAL, ENDODONTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL WAVEONE GOLD SMALL 3-FILE STER 31MM; FILE, PULP CANAL, ENDODONTIC Back to Search Results
Catalog Number A0753231G0S03
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.Additional information regarding the patient outcome has been requested and will be submitted as it becomes available.
 
Event or Problem Description
In this event it is reported that waveone gold small 3-file ster 31mm broke during use.The broken part could not be retrieved and patient was referred to endo specialist.
 
Additional Manufacturer Narrative
Involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1659896).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique (motor setting not communicated), overuse (like it may be the case here - file used 12 times while the instrument is a non-reusable / single use item), excessive wear, or material issue (no analysis of the broken fragments possible).Root causes are not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVEONE GOLD SMALL 3-FILE STER 31MM
Common Device Name
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key20462905
Report Number8031010-2024-00120
Device Sequence Number12732659
Product Code EKS
UDI-Device IdentifierD716WGSMALL311
UDI-PublicD716WGSMALL311
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 11/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA0753231G0S03
Device Lot Number1659896
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date10/04/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/16/2024
Supplement Report FDA Received Date11/27/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-