| Catalog Number |
A0753231G0S03 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.Additional information regarding the patient outcome has been requested and will be submitted as it becomes available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that waveone gold small 3-file ster 31mm broke during use.The broken part could not be retrieved and patient was referred to endo specialist.
|
| |
|
Additional Manufacturer Narrative
|
|
Involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1659896).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique (motor setting not communicated), overuse (like it may be the case here - file used 12 times while the instrument is a non-reusable / single use item), excessive wear, or material issue (no analysis of the broken fragments possible).Root causes are not identified.
|
| |
|