| Model Number |
DVBC3D4 |
| Medical Device Problem Codes |
Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
09/27/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented to the emergency department after receiving a shock.A remote monitoring transmission indicated that the patient was treated with shocks and anti-tachycardia pacing (atp) for an episode of ventricular tachycardia (vt) but the rhythm was suspected to be supra ventricular tachycardia (svt).The icd remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory had an observation relating to vt detection.Analysis of the device memory indicated unexpected delivery of ventricular tachyarrhythmia therapy.The device memory indicated an observation related to anti-tachycardia pacing therapy.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|