| Catalog Number |
91-9509SC / 519250 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Erosion (1750)
|
| Date of Event |
08/23/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, non-conforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no non-conforming reports or capas that were confirmed to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, the device was explanted and replaced due to cylinder migration and possible erosion.
|
| |
|
Event or Problem Description
|
|
According to the available information, the device was explanted and replaced due to cylinder migration and possible erosion.
|
| |
|
Additional Manufacturer Narrative
|
|
One malleable rod was received for evaluation.Visual examination of the returned components revealed no abnormalities that would have contributed to the report of migration and erosion.However, because examination of the components may not conclusively confirm or disprove the report of migration/erosion, quality accepts the physician¿s observations as to the reason for surgical intervention.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|