Model Number 3.0MM KNOTLESS SUTURETAK
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/03/2024
Type of Reportable Event Malfunction
Event or Problem Description
On 10/4/2024, it was reported by a sales representative via email that two ar-8934bck 3.0mm knotless suturetak anchors broke during insertion.This was discovered during a deltoid repair procedure on (b)(6) 2024.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: b5, g3, h3, h6.Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
Additional information received on 10/9/2024: all the broken fragments were removed.The case was completed using an ar-7220 fiberwire.The case was delayed fifteen minutes without additional anesthesia.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.The complaint allegation is confirmed.Photographic evidence provided from the field of an unpackaged ar-8934bck, with batch number 14946099, revealed a damaged anchor.Therefore, arthrex was able to deduce the cause of the complaint as misuse.The most likely cause for the reported failure can be attributed to misaligned insertion and/or prying/leveraging the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.0MM KNOTLESS SUTURETAK
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key20463686
Report Number1220246-2024-08169
Device Sequence Number7880801
Product Code MAI
UDI-Device Identifier00888867380981
UDI-Public(01)00888867380981(11)220401(17)260331(10)14946099
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K180594
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model Number3.0MM KNOTLESS SUTURETAK
Device Catalogue NumberAR-8934BCK
Device Lot Number14946099
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2024
Supplement Date Received by Manufacturer10/09/2024
08/15/2025
Initial Report FDA Received Date10/16/2024
Supplement Report FDA Received Date10/25/2024
08/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-