This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Correction is being made to previously submitted g3 - date received by manufacturer.The correct date was (b)(6) 2024.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the results of the investigation, it was not possible to determine the cause of the occurrence from the information obtained, it is possible that it was caused by physical stress from bumping the distal end of the scope or dropping the distal end, or chemical stress from the chemical solution used.However, the definitive root cause of the suggested event could not be determined.The event can be prevented by following the instructions for use (ifu): operation manual for ltf-190-10-3d important information ¿ please read before use ¿ dangers, warnings, and cautions: do not strike, hit, or drop the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.¿ reprocessing manual for ltf-190-10-3d 3.1 compatibility summary methods listed as ¿compatible¿ in table 3.1 and table 3.2 are compatible for routine use only when used according to manufacturer¿s instructions.Repeated use and reprocessing of endoscopes and accessories lead to gradual wear and tear.Furthermore, reprocessing methods that employ higher temperatures and more caustic/corrosive materials may lead to faster deterioration.In general, sterilization processes are harsher on equipment than disinfection processes.Before each patient procedure, inspect the endoscope and accessories for damage, according to the instructions described in this manual and its companion ¿operation manual¿.¿ 3.1 compatibility summary: list of compatible methods validated in terms of biological efficacy and material durability sterilization by sterrad® 50/100s/200/nx¿ system may deteriorate the adhesive of the insertion section.Depending on the circumstances, replacement of the insertion section may be required.Before use, confirm that the endoscope is free from any damage or other irregularities.For further details, contact olympus.Olympus will continue to monitor field performance for this device.
|