As reported, the tip of a cook® single-use holmium laser fiber broke off during a ureteroscopy procedure.A cook medical, ngage stone extractor was used to retrieve the fiber tip from the patient.A second laser fiber was used to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Investigation evaluation: description of event: as reported, the tip of a cook® single-use holmium laser fiber broke off during a ureteroscopy procedure.A cook medical, ngage stone extractor was used to retrieve the fiber tip from the patient.A second laser fiber was used to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), and quality control (qc) procedures of the device were conducted during the investigation.A device failure analysis was not conducted as the device was not returned by the customer.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that that the device was manufactured to specification and there is no evidence that nonconforming product exists in house or in the field.Cook also reviewed product labeling.The product ifu, contains the following information related to the reported failure mode.¿warnings¿do not bend fiber at sharp angles.¿ ¿precautions ¿never subject fiberoptics to sharp bends in handling, use, or storage.¿ ¿how supplied.¿ upon removal from package, inspect the product to ensure no damage has occurred.¿ cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|