NEOCIS INC. NEOCIS GUIDANCE SYSTEM; DENTAL NAVIGATION SYSTEM Back to Search Results
Model Number DRW-1000-00
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2024
Type of Reportable Event Serious Injury
Event or Problem Description
After a guided procedure, it was found that the implant placement was buccal compared to the surgical plan.A post-operative assessment with the system's log files was completed and a system accuracy check was completed.A definitive root cause for the inaccuracy reported could not be determined, there was no indication of a device malfunction based on investigation.The surgeon removed the implant and replaced it by hand.No further medical intervention was reported as a result of this issue.The handpiece is a standard 3rd party instrument used with yomi not manufactured by neocis.
 
Additional Manufacturer Narrative
Udi related data quality updates only.Updates based on udi/mdr data quality notification received on 13th dec 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOCIS GUIDANCE SYSTEM
Common Device Name
DENTAL NAVIGATION SYSTEM
Manufacturer (Section D)
NEOCIS INC.
545 nw 26th st
unit 700
miami FL 33127
Manufacturer (Section G)
NEOCIS INC.
545 nw 26th st
unit 700
miami FL 33127
Manufacturer Contact
alon mozes
545 nw 26th st
unit 700
miami, FL 33127
MDR Report Key20465156
Report Number3012787974-2024-80090
Device Sequence Number20781997
Product Code PLV
UDI-Device Identifier00810004900004
UDI-Public00810004900004
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K161399
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Dentist
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDRW-1000-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2024
Supplement Date Received by Manufacturer10/17/2024
Not provided
Initial Report FDA Received Date10/16/2024
Supplement Report FDA Received Date10/29/2024
04/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/06/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-