|
Reportable based on the device analysis completed on 1 oct 2024.It was reported that balloon pinhole at the distal markerband occurred.The 75% stenosed target lesion was located in the moderately tortuous vessel.A 3.5-4/4t/90 symmetry balloon catheter was advanced for dilatation.During the procedure, initial inflation pressure didn't change.The deflator indicated the balloon caused the issue.Instead of rupturing, pressure leakage seems to be the problem.The device was removed without any problem, and the procedure was completed with another of the same device.There were no patient complications reported, and the patient was in good condition after the procedure.However, product analysis revealed pinhole located approximately at the distal markerband.
|