H6: health effect term/code not available: suggested code : failure to resorb.An investigation was performed: no initial evaluation on the device could be performed because there was no returned sample for this complaint.The sample was not returned for this complaint and there was limited patient-related information provided for this event; therefore, a thorough investigation was not able to be performed.Investigation into several of the harms reported in this complaint was performed from a clinical literature review, as well as an investigation into the design and risk of the biozorb product.The harms identified during this complaint are: major pain (pain/discomfort/device palpability/sleep disfunction/failure to resorb), seroma, infection, major additional intervention (additional surgery/device explantation/calcifications).A review of clinical literature performed on lumpectomies and their associated complications was performed.A summary of this review concluded that, while lumpectomy is a well-established option for breast cancer treatment, it inherently carries risks of complications, most notably infections, seroma formation, chronic pain, and wound healing issues.The literature demonstrates variation in complication rates based on surgical technique, additional therapies (such as intraoperative radiation therapy (iort) or radiation), and individual patient factors, with infection rates ranging from 2% to 10% and seroma rates from 11% to 65%.Chronic pain, particularly neuropathic pain, affects nearly one-third of patients.(kwee et al.(2024)), migration can be influenced by several patient related factors which can influence the likelihood of migration after lumpectomy, such as breast density, biopsy location, and biopsy approach.Per a study by ashali jain et al.(2017), marker migration is a relatively common occurrence following stereotactic biopsies, happening in 13% of cases.Swelling is one of the expected events that could occur after surgery and/or radiation therapy regardless of the use of a biozorb device.Due to this, confirmation of the swelling during this event as a normal course of surgery vs being caused by the biozorb is not able to be done.Device explantation is based on the patient¿s physician¿s evaluation of their physical and health state and recommendation; therefore, an investigation into this intervention is not able to be performed for this complaint.Though complications are relatively common, most are manageable with appropriate postoperative care, making lumpectomy a viable option for many breast cancer patients when carefully monitored and followed up.An investigation was performed by subject matter experts (sme¿s) into several of the harms identified as being experienced by the patient related to design, risk management, and clinical factors : based on the clinical literature review, the harms reported to be experienced by the patient in the complaint are common symptoms from surgery and lumpectomies; therefore, a confirmation of the biozorb causing the harms was not able to be done.However, based on the review of the top harms in regards to the biozorb design and risk documentation, the adverse events reported can be tracked back to a potential shared cause: lack of adequate user needs for the product.User needs define the framework for product development and lay the ground for the definition of design inputs, design outputs and later verification and validation activities.Other potential causes are related to inadequate test methods and/or the lack of evidence to support some specific product requirements.When comparing the safety outcomes of the biozorb marker with general lumpectomy complications, it becomes evident that some of the complications associated with biozorb may stem from the lumpectomy procedure itself.Both procedures share overlapping risks, although the biozorb device introduces additional considerations.However, it has been confirmed that the lack of defined user needs and inadequate test methods and testing evidence are potential root causes for the harms reported in this complaint.This cause will be further investigated and addressed in a capa.The risks associated with this complaint event are further described and evaluated against the product and its hazard analysis in the health risk assessment (hra) biozorb complaint analysis.Future events will be monitored and trended.A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.
|