• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC SUREFORM; STAPLER 60 RELOAD GREEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC SUREFORM; STAPLER 60 RELOAD GREEN Back to Search Results
Model Number 48360G-09
Medical Device Problem Codes Failure to Form Staple (2579); Detachment of Device or Device Component (2907); Firing Problem (4011)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/20/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a da vinci-assisted low anterior resection surgical procedure, there was a problem with stapling the colon using the sureform 45 and 60 staplers with the sureform green 45 and 60 reloads.Stapling of the colon was not successful, and charging stopped halfway, preventing normal use of the device.There was a message on the da vinci console noting, "tissue too thick for stapling" despite normal tissue size.There was no patient consequence, but they noted that it was a very complicated situation at the time.Prior to calling technical support, the site already tried to use two instruments with blue and black reloads.The reporter stated that the surgeon was experienced with this kind of instruments/use.When the surgeon was using the stapler, both stapling and cutting ended mid-course, stated the caller.The intuitive surgical, inc.(isi) technical support engineer (tse) found nothing relevant in the logs.The tse asked the customer to reseat the sterile adapter (sa) prior to retrying with the sureform stapler, but the behavior remained.The caller stated that the robotic part of the procedure was now completed.The procedure was continuing as planned.Intuitive surgical, inc.(isi) followed up with the site and obtained the following additional information: the reported event involved sureform stapler 60 (p/n: 480460 / lot: 82230731) and sureform staple5 45 (p/n: 480445 / lot: t18230918).The stapler reload colors that were used throughout the procedure were blue, green, and black.The surgeon selected the black reload due to tissue thickness, although the tissue did not appear overly thick.Several fires occurred in a row, even at the end, and the stapler worked only partially.The surgeon was stapling the lower rectum, targeting non-calcified tissue that had been exposed to radiation or chemotherapy, with no tissue tension or bunching.The issue involved partial fires, with no tissue being pushed forward or jaws opening unexpectedly.There were no malformed staples, exposed blades, missing staples, or gaps in the staple line, but it was unclear if the reload was stuck to the tissue.No obstructions were encountered, and the surgeon may have fired over an existing staple line.Buttress material was used, and no clamping issues or bleeding were observed.Five reloads were needed to complete the stapling, and the procedure was completed.The reloads and instruments are not available for return, but photos and videos were provided for review.The surgeon described the situation as unpleasant, with the stapler jamming repeatedly and indicating tissue thickness issues, despite the tissue not being overly thick.Ultimately, a good anastomosis was achieved.
 
Additional Manufacturer Narrative
A return material authorization (rma) was issued to evaluate the sureform 60 green reload; however, the customer reported that the reload is not available for return.A review of an image provided by the customer was consistent with staples not releasing from the reload pocket.The root cause of the failure mode cannot be confirmed without the returned device.A review of the videos provided by the customer was performed by an intuitive surgical, inc.(isi) clinical development engineer (cde).The review of the videos showed unformed and malformed staples within the patient.It also showed a fenestrated bipolar forceps (fbf) instrument had an unformed staple in its jaw and then the jaw was moved further back into the patient's cavity.When the jaw of the fbf reappeared, the staple was no longer there.It is unknown if the staples were retrieved from the patient¿s anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFORM
Common Device Name
STAPLER 60 RELOAD GREEN
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
4085232100
MDR Report Key20468731
Report Number2955842-2024-20621
Device Sequence Number7228370
Product Code GDW
UDI-Device Identifier10886874113841
UDI-Public(01)10886874113841(10)K10231005
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K173721
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48360G-09
Device Catalogue Number48360G
Device Lot NumberK10231005
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/20/2024
Initial Report FDA Received Date10/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient SexMale
-
-