| Model Number |
A22043A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the ceramic tip of the rigid endoscope sheath was broken.It is unknown when the issue occurred.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The customer's allegation was confirmed.Based on the results of the investigation, it is likely that the following led to the malfunction: thermally and mechanically induced damage.It is not possible to say whether there was prior damage or wear to the insulating insert, whether the damage was triggered during the last reprocessing (cds) of the instrument, or during the last procedure.A definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
The issue was found during reprocessing.
|
| |
|