Event or Problem Description
|
|
The recipient came with his family to our distributor since he reported that his hearing performance with the device is affected in the right ear.Re-implantation is considered.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The recipient came with his family to our distributor since he reported that his hearing performance with the device was affected in the right ear.Re-implantation is considered.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress seems very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The recipient came with his family to our distributor since he reported that his hearing performance with the device was affected in the right ear.The recipient has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|