| Catalog Number |
8065752066 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
The physician reported that product could not be attached to a ultrasound handpiece during a test.The surgery was completed after replacing the product with another one.There was no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in d.4., h.3., h.6.And h.11.One opened phaco tip with a wrench, in a bag, in a box was received for the report.Sample was visually inspected and found to be conforming.Functional thread check was performed and found to be conforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned sample was found to be conforming for visual and functional testing associated with the reported event; therefore, product could not be attached to ultrasound (us) handpiece was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the phaco tip was manufactured to specifications.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|