|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information added: corrected d5, added: h3, h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the image sensor unit may have been broken, leading to the subject event.The probable cause of breakage may be irregular stress to the bending section, leading to the event since multiple damaged sites were observed on the bending section.However, the cause of it was unable to be specified.A definitive root cause could not be determined.The subject event can be detected by implementing in accordance with the below ifu.Instructions for ltf-vh operation manual chapter 3, ¿preparation and inspection¿ section 3.6, ¿inspection of the endoscopic system¿ ¿¦inspection of the endoscopic image¿ olympus will continue to monitor field performance for this device.
|