WRIGHT MEDICAL TECHNOLOGY INC INFINITY POLY SZ 3 6MM TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 33653306
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Osteolysis (2377); Inadequate Osseointegration (2646); Subluxation (4525); Joint Impingement (4834)
Date of Event 04/07/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer received a report from the national joint registry that contains unpublished collected data on the usage and the outcomes of the ankle joint replacement.The report details analysis provided for revision procedures performed between january 2018 ¿ december 2023.During the review of the report, it was identified that on (b)(6) 2021 a patient required revision surgery due to aseptic loosening & component migration/ dissociation, which was not previously reported to the manufacturer.
 
Additional Manufacturer Narrative
The reported event of a patient requiring revision surgery due to aseptic loosening and component migration or dissociation could not be confirmed, since the device was not returned for evaluation and no other additional information will be made available by the national joint registry.More detailed information about the patient's medical history, the event details and the involved device(s) must be available to determine the root cause.If any additional information becomes available, the investigation will be reopened and re-evaluated accordingly.Please note, that reports received from the national joint registry are not published reports and therefore web link is not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY POLY SZ 3 6MM TOTAL ANKLE SYSTEM
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key20472615
Report Number3010667733-2024-00761
Device Sequence Number7603283
Product Code HSN
UDI-Device Identifier00840420122704
UDI-Public00840420122704
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K140749
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number33653306
Device Lot Number1645490
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2024
Initial Report FDA Received Date10/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight61 KG
-
-