• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VISIONSENSE LTD. VS3 IRIDIUM - LIGHT INTEGRATOR (BEAMCOMBINER); CONFOCAL OPTICAL IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VISIONSENSE LTD. VS3 IRIDIUM - LIGHT INTEGRATOR (BEAMCOMBINER); CONFOCAL OPTICAL IMAGING Back to Search Results
Model Number 161-0001
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2021
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during use, the unit laser had no output from beam combiner- fiber pinched.There was no patient injury.
 
Additional Manufacturer Narrative
This report is being submitted as part of remediation activities associated with capa # (b)(4), which is addressing mdr reporting gu idance from fdas 1995 preamble, which ensures complaints related to corrections and recalls which were previously reported to fda under 21 cfr 806 are now reported as mdrs; this is not a new malfunction/event.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had a cable/wire failure.It was reported that the light was not clear or was dim.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur due to mishandling or improper transportation.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VS3 IRIDIUM - LIGHT INTEGRATOR (BEAMCOMBINER)
Common Device Name
CONFOCAL OPTICAL IMAGING
Manufacturer (Section D)
VISIONSENSE LTD.
20 hamagshimim st
petach-tikva
IS 
Manufacturer (Section G)
VISIONSENSE LTD.
20 hamagshimim st
petach-tikva
IS  
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20473928
Report Number9616637-2024-00008
Device Sequence Number7228531
Product Code OWN
UDI-Device Identifier10884521817555
UDI-Public10884521817555
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K150018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 10/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number161-0001
Device Catalogue Number161-0001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2022
Initial Report FDA Received Date10/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0283-2023
Patient Sequence Number1
Patient SexUnknown
-
-