Manufacturer's investigation conclusion: the centrimag console was evaluated following the reported event of the centrimag motor not being able to be recognized by the console.There were no photos or log files submitted for review.The centrimag console was not returned; however, the centrimag motor was returned and evaluated at the service depot where the reported event was confirmed.Continuity testing revealed an open line on the b2+/b2- line and high resistance on the hx/hy signal line.The entire motor cable was dissected and the bearing phase b2 line (brown power wire) was found to be open.The cable was stripped and slight tugging near the proximal bend relief resulted in the brown cable being severed indicating there was severe wire fatigue.This issue is covered under the centrimag motor investigation.Multiple attempts were made to obtain additional information from the customer regarding the serial number of the console; however, no response was received.The reported event was not correlated to the cenrimag console.The patient handbook cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the 2nd generation centrimag system operating manual section 12.1 entitled "appendix i ¿ console alarms and alerts" contains a list of console alarms and alerts, including motor-related alarms, as well as appropriate operator response to these events.Device history records could not be reviewed due to the serial number of the centrimag console being unavailable.No further information was provided.The manufacturer is closing the file on this event.
|