AVANOS MEDICAL INC. GASTROINTESTINAL ANCHOR SET, SAF-T-PEXY T-FASTENERS; DH EF PERC PLACEMENT PRODUCTS
|
Back to Search Results |
|
| Model Number |
98701 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported three gastropexy anchors had cracks noted during procedure.Additional information received 26-sep-2024 noted "the kit that malfunctioned was used and we placed one of the anchor set.".
|
| |
|
Additional Manufacturer Narrative
|
|
The actual complaint product was returned for evaluation.A review of the device history record is not possible as no lot number was provided; therefore, the udi-pi is unavailable.The lot number was not provided by the customer.All information reasonably known as of 17-oct-2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The sample device (s) were evaluated.The needles were each in the protective cover.The suture and anchor for sample one was attached and have not been deployed.There were no obvious cracks or breaks on the hubs.Using a syringe, water was infused through the needle hubs.The water leaked out on two sides of the hubs.The area was examined under magnification.There was a hole on two opposing sides of the clear hub, just under the luer connector section.There was also a crack with stress whitening on one side by the hole.The mold number was "1".The suture and anchor for sample two was attached and have not been deployed.There were no obvious cracks or breaks on the hubs.Using a syringe, water was infused through the needle hubs.The water leaked out on two sides of the hubs.The area was examined under magnification.There was a hole on two opposing sides of the clear hub, just under the luer connector section.The mold number was "1".The subject sample device(s) provided were evaluated confirming a hole on two opposing sides of the clear hub, just under the luer connector section; however, process assessment found no activities or tools that could cause this type of flaw in the gastropexy component.Root cause could not be determined.All information reasonably known as of 13-nov-2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|