ONKOS SURGICAL ELEOS LIMB SALVAGE SYSTEM; TIBIAL BASEPLATE Back to Search Results
Lot Number 93643
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 09/18/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by (b)(4), an onkos distributor, that a 43-year-old female patient with an eleos distal femur replacement underwent a revision due to loosening of the eleos cemented segmental stem that was inserted into the femur without the use of bone cement.The reported information also indicated that the eleos tibial baseplate was also inserted without bone cement and the surgeon suspected that the patient had a peri-operative infection.
 
Additional Manufacturer Narrative
Based on the review of device history records, the investigation concluded that the root cause of the reported loosening and perioperative joint infection was not related to the design, manufacture, and/or sterilization of the revised eleos devices.Based on the reported information, the most probable root cause of the femoral cemented segmental stem loosening is the failure to use bone cement to fixate the device as detailed in the instructions for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS LIMB SALVAGE SYSTEM
Common Device Name
TIBIAL BASEPLATE
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer Contact
yara neurauter
77 east halsey road
parsippany, NJ 07054
8447672766
MDR Report Key20475706
Report Number3013450937-2024-00324
Device Sequence Number17950976
Product Code KRO
UDI-Device IdentifierB278TB2202E01M0
UDI-PublicB278TB2202E01M0
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Lot Number93643
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2024
Initial Report FDA Received Date10/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AP DISTAL FEMUR AXIAL PIN; DF DISTAL FEMUR; HC CEMENTED MODULAR SEGMENTAL STEM; HR MODULAR COLLAR LOCKING RING; PB MODULAR PLASMA COLLAR; PS TIBIAL POLY SPACER; TB TIBIAL BASEPLATE; TH TIBIAL ROTATIONAL HINGE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient SexFemale
-
-