H3: code "other" was selected as the medical device was discarded at facility.Return not possible.H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.According to the gore® dryseal flex introducer sheath instructions for use (ifu), potential device or procedure-related adverse events that may occur and/or require intervention include, but are not limited to: vascular trauma w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
On (b)(6) 2024, this patient underwent endovascular treatment of an abdominal aortic aneurysm.A gore® dryseal flex introducer sheath (dsf sheath) was used for access.The dsf sheath was inserted from right side.It was reported percutaneous transluminal angioplasty(pta) was performed from right common iliac artery down to the right external iliac artery due to the vessel was narrow, but the physician still felt a resistance while advancing the dsf sheath around the right external iliac artery.After all stent grafts were placed, a dissection on the right external iliac artery was revealed on an access angiography.A bare stent was placed to treat the dissection.The patient tolerated the procedure.
|