EXACTECH, INC. LOGIC CR FEMORAL POR, LEFT, SZ 5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-010-04-0250 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H3: the revision reported may have been the result of prosthesis wear and femoral loosening.The wear and/or deformation of the tibial insert and patella may have been the result of third body wear, orientation of the components, patient related conditions, instability, or any combination of these possibilities.The loosening may have been the result of an insufficient bond between the implant and the bone which led to aseptic (non-infected) loosening of the femoral component.However, this cannot be confirmed because the components were not returned for evaluation and radiographs were not provided.
|
| |
|
Event or Problem Description
|
|
As reported, approximately 10 months and 3 days post the previous revision surgery, the patient underwent a second revision due to patella wear and poly wear; there was no change in sizing.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.Associated with 1038671-2023-02948 and 1038671-2024-03169.Previous revision reported under 1038671-2023-00115.
|
| |
|