• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. PRUITT F3 CAROTID SHUNT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. PRUITT F3 CAROTID SHUNT Back to Search Results
Catalog Number E2012-10
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was discarded and not returned for evaluation.Therefore, the exact root cause of the balloon rupture could not be determined.As stated in the ifu: exercise caution when encountering extremely diseased vessels.Arterial rupture or balloon failure due to sharp calcified plaque may occur.The possibility of balloon rupture must be taken into account when considering the risks involved in the endarterectomy procedure.The lot history record for the device was reviewed, no issues were found during manufacturing or packaging that would cause or contribute to the reported event.We have not received any similar complaints related to this lot number.Note: this is report 1 of 2 related to the first event/device.Report 1220948-2024-00211 was submitted for the second device/event.
 
Event or Problem Description
It was reported that the pruitt f3 carotid shunt balloon ruptured during a carotid endarterectomy procedure.The issue was detected due to the balloon leaking and was clamped to address it.No injury was reported to the patient.Note: this is report 1 of 2 related to the first event/device.Report 1220948-2024-00211 was submitted for the second device/event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRUITT F3 CAROTID SHUNT
Common Device Name
CAROTID SHUNT
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC
63 second ave
burlington MA 01803
Manufacturer Contact
peter song
63 second ave
burlington, MA 01803
7814251683
MDR Report Key20481709
Report Number1220948-2024-00205
Device Sequence Number12863428
Product Code MJN
UDI-Device Identifier00840663101283
UDI-Public(01)00840663101283(11)240624(17)290528(10)XPF00001235(91)E2012-10
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K182916
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 10/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberE2012-10
Device Lot NumberXPF00001235
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2024
Initial Report FDA Received Date10/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/24/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-