EXACTECH, INC. TRULIANT TIB FIT TRAY CEM SZ 4F / 4T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-022-45-4040 |
| Medical Device Problem Codes |
Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Pain (1994); Discomfort (2330); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Joint Laxity (4526); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported via legal documentation the patient had a left knee revision on (b)(6) 2020.Approximately 2 years after the first revision the patient had a second left knee revision on (b)(6) 2022 due to accelerated and preventable wear.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
1038671-2023-02942 the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, loosening and/or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.
|
| |
|