C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO 4F SL W/GUARDIVA, PROBE COVER, & BELDAR NEEDLE GUIDES; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Swelling/ Edema (4577)
|
| Date of Event |
08/30/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer that patient came in with a sl picc placed on 7/22.The picc was out 2.5 cm on arrival.When flushing, saline came out the insertion site.On removal, picc was slit at the 4cm mark.Patient denies any damage to the catheter or issues with flushing previously.Very similar symptoms to the other picc we had with this problem.Picc is not in the same lot #.Slit was 1-1.5cm beneath the skin.The impact to the patient was mild swelling to arm and having to have a new picc line placed (unnecessary procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak was confirmed.The product returned for evaluation was one 4 fr sl powerpicc solo catheter.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue occurs due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated and a split was observed in the catheter tubing between the 3 cm and 4 cm depth markers.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: damage which was circumferentially aligned.Fracture edges which were rounded and polished due to repeated material wear.'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing) repetitive kinking of the indwelling catheter appears to have contributed to the observed leak; however, the specific circumstances surrounding how kinking developed into a fracture within the catheter tubing were ultimately unknown.A measurement of the catheter outer diameter, inner diameter and wall thickness were taken.The catheter was found to be within specification.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer that patient came in with a sl picc placed on 7/22.The picc was out 2.5 cm on arrival.When flushing, saline came out the insertion site.On removal, picc was slit at the 4cm mark.Patient denies any damage to the catheter or issues with flushing previously.Very similar symptoms to the other picc we had with this problem.Picc is not in the same lot #.Slit was 1-1.5cm beneath the skin.The impact to the patient was mild swelling to arm and having to have a new picc line placed (unnecessary procedure.
|
| |
|