BAXTER HEALTHCARE CORPORATION POLYFLUX; DIALYZER, CAPILLARY, HOLLOW FIBER Back to Search Results
Catalog Number 103530
Medical Device Problem Codes Break (1069); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an external fluid leak was observed from the header of a polyflux 140h during hemodialysis therapy.The device was replaced to continue treatment.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Additional information: h6, h11.H11: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photo showed the device in use for therapy, and a water droplet is visible outside of the access header.The specific defect that allowed the leakage was not visible in the photo.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition was not determined as no further or actual sample testing could be performed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYFLUX
Common Device Name
DIALYZER, CAPILLARY, HOLLOW FIBER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
holger - crafoord - strasse 26
hechingen D-723 79
GM   D-72379
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key20482970
Report Number9611369-2024-00190
Device Sequence Number4208313
Product Code FJI
UDI-Device Identifier07332414037925
UDI-Public07332414037925
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K043342
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Catalogue Number103530
Device Lot Number3-5705-H-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2024
Supplement Date Received by Manufacturer10/29/2024
Initial Report FDA Received Date10/18/2024
Supplement Report FDA Received Date11/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-