Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.Additional devices included on this report are as follows: item # vbhr101002a/ lot # unk, gore® viabahn® endoprosthesis were overlapped and implanted in same anatomical location.Multiple attempts were made to obtain additional information from the physician.However, no further details were provided that would help us investigate further.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
The following information was reported to gore: on (b)(6) 2024, two gore® viabahn® endoprosthesis (vsx device) were delivered to treat a popliteal aneurysm.On (b)(6) 2024, the patient returned to hospital with thrombosed vsx devices.During this time the patient was diagnosed positive for heparin-induced thrombocytopenia (hit).It was reported that the patient did receive iv heparin at the time of the procedure as well.The patient underwent mechanical thrombectomy and lysis.The graft was relined with vsx devices without cbas® heparin surface lining.
|
|
Section h6 updated to reflect completion of investigation.A review of manufacturing records verified that the lot involved in this complaint met all pre-release specifications.The device remains implanted and was, therefore, not available for analysis.No clinical images enabling direct assessment of product performance were returned for evaluation.Cause of the reported event cannot be established based on the information reported to gore.Further information regarding this event was requested by gore, but no further information has been reported, therefore this investigation is considered complete.
|