MAUDE Adverse Event Report: ZIMMER DENTAL IMPL TWIST MP-1 3.75 MM 1 0 MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMPL TWIST MP-1 3.75 MM 1 0 MM; DENTAL IMPLANT Back to Search Results
Catalog Number 1989
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint (b)(4).G4: additional 510(k) number is k943604.
 
Event or Problem Description
It was reported that the mount won't come off.After the spline implant was inserted, the mounting screw was too tight to remove, so implant was removed.Another implant was inserted.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie complaint cmp-(b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem d4: additional device information d9: device availability g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h4: device manufacturer date h6: adverse event problem h10: additional narrative zimvie received one (1) 1989, (impl twist mp-1 3.75 mm 1 0 mm) for evaluation.Visual evaluation of the as returned devices identified the implant and bundle mount with signs of use.The implant's original packaging wasn¿t returned for evaluation.The implant was attached to the mount and the devices wouldn¿t disengage/release, during a functional / physical test.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number 2023030198.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2023030198 for similar events and no other complaint was identified.Review completed utilizing keywords: does not disengage/release.The customer did not submit images for the reported event.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are incorrect techniques used - customer error, excessive torque applied.Therefore, based on the available information, and functional testing a device malfunction did occur.The reported event/coob was non-verifiable since the condition of the product/packaging received by the customer was unknown / non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TWIST MP-1 3.75 MM 1 0 MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key20485990
Report Number0002023141-2024-03321
Device Sequence Number4106071
Product Code DZE
UDI-Device Identifier00889024013124
UDI-Public(01)00889024013124(17)280426(10)2023030198
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K013494
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1989
Device Lot Number2023030198
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/18/2024
Supplement Report FDA Received Date03/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-