Zimvie complaint cmp-(b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem d4: additional device information d9: device availability g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h4: device manufacturer date h6: adverse event problem h10: additional narrative zimvie received one (1) 1989, (impl twist mp-1 3.75 mm 1 0 mm) for evaluation.Visual evaluation of the as returned devices identified the implant and bundle mount with signs of use.The implant's original packaging wasn¿t returned for evaluation.The implant was attached to the mount and the devices wouldn¿t disengage/release, during a functional / physical test.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number 2023030198.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2023030198 for similar events and no other complaint was identified.Review completed utilizing keywords: does not disengage/release.The customer did not submit images for the reported event.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are incorrect techniques used - customer error, excessive torque applied.Therefore, based on the available information, and functional testing a device malfunction did occur.The reported event/coob was non-verifiable since the condition of the product/packaging received by the customer was unknown / non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|